Preview

Problems of Particularly Dangerous Infections

Advanced search

Construction and Approval of the Test-System for the Detection of Antibodies to Anthrax Agent Using Indirect Fluorescent Immunoassay

https://doi.org/10.21055/0370-1069-2013-4-76-78

Abstract

Constructed and validated has been the test-system for the detection of antibodies to anthrax agent using indirect fluorescent immunoassay. The panel consists of antigen preparation, positive control sample, negative control, and FITS-labeled rabbit antibodies. Studied have been 10 acapsular B. anthracis strains with different plasmid content and colony morphology. B. anthracis vaccine strain STI-1 is demonstrated to be the optimal one as antigen preparation. Sensitivity of this test-system is not less than 1:40, the working dilution of FITC-conjugate being 1:16. Specificity of the panel has been studied using 100 sera of healthy donors. Approved is 100 % reproducibility of the technique at different time intervals, as well as when carried by different specialists. The test-system allows for confirmation of anthrax diagnosis in humans.

About the Authors

L. Yu. Aksenova
Stavropol Research Anti-Plague Institute
Russian Federation


A. G. Ryazanova
Stavropol Research Anti-Plague Institute
Russian Federation


E. V. Zhdanova
Stavropol Research Anti-Plague Institute
Russian Federation


O. I. Tsygankova
Stavropol Research Anti-Plague Institute
Russian Federation


N. P. Buravtseva
Stavropol Research Anti-Plague Institute
Russian Federation


E. I. Eremenko
Stavropol Research Anti-Plague Institute
Russian Federation


I. S. Tyumentseva
Stavropol Research Anti-Plague Institute
Russian Federation


O. I. Kogotkova
Stavropol Research Anti-Plague Institute
Russian Federation


A. A. Zuenko
Stavropol Research Anti-Plague Institute
Russian Federation


A. N. Kulichenko
Stavropol Research Anti-Plague Institute
Russian Federation


References

1. Ryazanova A.G., Eremenko E.I., Tsygankova E.A., Aksenova L.Yu., Tsygankova O.I., Buravtseva N.P., Golovinskaya T.M., Kulichenko A.N. [Analysis of the anthrax morbidity rate in the Russian Federation in 2011, and prognosis for 2012]. Probl. Osobo Opasn. Infek. 2012; 1(111): 37–9.

2. Marinin L.I., Onishchenko G.G., Kravchenko T.B., Dyatlov I.A., Tyurin E.A., Stepanov A.V., Nikiforov V.V. [Human Anthrax: Epidemiology, Prophylaxis, Diagnostics, and Treatment]. M.: ZAO MP “Gigiena”; 2008. 416 p.

3. Tyumentseva I.S., Afanas’ev E.N., Efremenko V.I., Bazikov I.A., Alieva E.V. [Method of diagnostic serum production]. RF Patent 213510.


Review

For citations:


Aksenova L.Yu., Ryazanova A.G., Zhdanova E.V., Tsygankova O.I., Buravtseva N.P., Eremenko E.I., Tyumentseva I.S., Kogotkova O.I., Zuenko A.A., Kulichenko A.N. Construction and Approval of the Test-System for the Detection of Antibodies to Anthrax Agent Using Indirect Fluorescent Immunoassay. Problems of Particularly Dangerous Infections. 2013;(4):76-78. (In Russ.) https://doi.org/10.21055/0370-1069-2013-4-76-78

Views: 802


Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 License.


ISSN 0370-1069 (Print)
ISSN 2658-719X (Online)